Steps: Connect the Reader (starting firmware version for ex: 2.2.9) 1. Launch the application 2. Click Next on the Welcome screen. 3. Disconnect the Reader Expected Result: The Next button is disabled in the Ready to Start Update screen after disconnect the reader as per UI flow. Actual Result - Next button is still enabled after disconnecting the Reader Build Tested: LibreUpdate_744_161 (both windows and mac)
the expected behavior from UI flow page 1. the text between 'connection Reader Screen' and 'Ready to Update Screen' mentions expected behavior. "Note: If the reader is disconnected at this prompt, then the “next” button will be inactivated." with current behavior, it would fail the SIT test protocol.
Fixed in build 756
Verified the issue in the build LibreUpdate_765_158 and still the issue persists on both Windows and Mac.
Fixed in build 770
Created attachment 29 [details] Bug 55
Verified in the build 778 on both windows and mac and working as expected. Screen shot attached.