Bug 55

Summary: SIT_TC3_TS3 - Next Button is enabled after disconnecting the Reader in Ready to Start Update screen
Product: TestProduct Reporter: Mohan Kumar Karruppusamy <mohankumar.karruppusamy>
Component: TestComponentAssignee: Brian Roberge <brian.roberge>
Status: VERIFIED FIXED    
Severity: critical CC: kuanshen.an
Priority: Highest    
Version: unspecified   
Hardware: All   
OS: All   
Attachments: Bug 55

Description Mohan Kumar Karruppusamy 2018-08-08 21:20:04 EDT
Steps:

Connect the Reader (starting firmware version for ex: 2.2.9)

1. Launch the application
2. Click Next on the Welcome screen.
3. Disconnect the Reader

Expected Result: The Next button is disabled in the Ready to Start Update screen after disconnect the reader as per UI flow.

Actual Result - Next button is still enabled after disconnecting the Reader

Build Tested: LibreUpdate_744_161 (both windows and mac)
Comment 1 Kuanshen An 2018-08-09 16:44:13 EDT
the expected behavior from UI flow  page 1. the text between 'connection Reader Screen' and 'Ready to Update Screen' mentions expected behavior.

"Note: If the reader is
disconnected at this
prompt, then the “next”
button will be inactivated."

with current behavior, it  would fail the SIT test protocol.
Comment 2 Brian Roberge 2018-08-14 14:03:03 EDT
Fixed in build 756
Comment 3 Mohan Kumar Karruppusamy 2018-08-15 18:50:14 EDT
Verified the issue in the build LibreUpdate_765_158 and still the issue persists on both Windows and Mac.
Comment 4 Brian Roberge 2018-08-15 20:14:29 EDT
Fixed in build 770
Comment 5 Mohan Kumar Karruppusamy 2018-08-17 23:57:53 EDT
Created attachment 29 [details]
Bug 55
Comment 6 Mohan Kumar Karruppusamy 2018-08-17 23:59:08 EDT
Verified in the build 778 on both windows and mac and working as expected.

Screen shot attached.