| Summary: | RSU app is hung when switching the reader before software upgrade progress bar starts | ||
|---|---|---|---|
| Product: | TestProduct | Reporter: | Kirankumar Shiva <kirankumar.shiva> |
| Component: | TestComponent | Assignee: | No one <sysadmin> |
| Status: | VERIFIED FIXED | ||
| Severity: | minor | CC: | brian.roberge |
| Priority: | Highest | ||
| Version: | unspecified | ||
| Hardware: | PC | ||
| OS: | All | ||
| Attachments: | Issue Log | ||
Fixed in build 1028 Verified. Build Details: build : 1028 Installer: 154 Reader F/W : 2.1.2 |
Created attachment 83 [details] Issue Log Build details : Build : 1022 Installer : 157 OS : Mac and Windows Readers FW : 2.2.13 Steps to reproduce : 1) Connect the reader to RSU installed PC. 2) Launch the Application and navigate to Ready to Start Update screen 3) Click Next button on the Ready to Start Update screen and immediately disconnect Reader 1 and connect Reader 2 before progress bar appears. Actual : RSU app is hung and "update unsuccessful" is not displayed. Expected : Update Unsuccessful Screen with error message "Update Unsuccessful" is displayed. Note : Reproducible in MacOS. Attached the log